Product Information

P75 Venous Pressure Transducer

Product Name: P75 Venous Pressure Transducer
Product Type:
Exhibitors: HSE
Product Documentation: No Related Documents

Brief

Venous Pressue Transducer used in PLUGSYS system


P75 Venous Pressure Transducer  Detailed description
The P75 has a removable Macrolon® dome with a pressure connection and a vent connection at the side, so that it can be filled free of air bubbles. The dome connections have a male Luer taper so that suitable stopcocks* can be attached. The transducer has a metal housing. The actual pressure sensor inside is made from ceramic and therefore has excellent resistance to different media. The transducer’s rugged construction can withstand pressure overloads up to 4000 mmHg without damage. It works together with any DC bridge amplifier (e.g., PLUGSYS TAM-A).
Specifications
Pressure Range ±75 mmHg (±100 cmH2O)
Volume Displacement 0.06 mm3 /10 mmHg
Linearity ±0.15 mmHg
Long-Term Drift ±0.04 mmHg
Overload -760 (= vacuum) to 4000 mmHg
Measurement Media All gases and liquids which do not attack Macrolon®
Temperature Range 0° to 50°C
Zero Drift ±0.04 mmHg/10°C (0° to 50°C)
Range Drift ±0.04 mV/10°C (±0.04 mmHg/10°C) (0° to 50°C)
Electrical Data:
Supply Voltage 5 V (4.5 to 5.5 V) DC only
Current Loading 15 mA max.
Sensitivity 1 mV/mmHg, nominal
Output Resistance 300 Ω, nominal
Frequency Range 0 to 100 Hz
Connection Cable Approx. 1.5 m (4.9 ft) long
Suitable Amplifiers Any bridge amp providing 5V DC excitation voltage